如需购买服务产品,请与思柯拉特联系
Call Hotline:(+86)18151568323
Product Identifiction |
TrueRay 350 9962800 |
Operating environment conditions | |
---|---|
Ambient temperature for operation | 10℃ ~ 40℃ |
Relative humidity for operation | 30% ~ 75% |
Atmospheric pressure for operation | 700 hPa ~ 1.060 hPa |
Cooling method | Oil-immersion natural cooling, insulation strength ≥ 35 kV / 2.5mm, oil temperature 10℃ ~ 55℃ |
Power supply | The high voltage generator used by medical X-ray computed tomography equipment (CT) is used as a specific power supply. |
Nominal X-ray tube voltage | 140KV |
|
|
Filament voltage characteristics | |
Focus nominal value |
Small focus F1:0.8X0.4 Large focus F2: 0.8 X 0.7 |
Anode target diameter | 140mm±0.5mm |
Anode target angle | 8.0°±0.25° |
Weight | 8.6kg±0.3kg |
Safety classification | GB 9706.228-2020 |