TrueRay Chorus

Applicable CT Models
Somatom go.Now
go.Up

Technical Parameter
Product Identifiction
TrueRay Chorus
9962800
Operating environment conditions
Ambient temperature for operation 10 ℃~40 ℃
Relative humidity for operation 30%~75 %
Atmospheric pressure for operation 700 hPa ~ 1.060 hPa
Cooling method Oil-immersion natural cooling, insulation strength ≥ 35 kV / 2.5mm, oil temperature 10℃ ~ 55℃
Power supply Power supply complies with the GB/T 12079-2012 regulations. Source power 50 kW, frequency ≤ 150 Hz
Nominal X-ray tube voltage 140KV

Filament voltage characteristics
Nominal anode input power Small focus F1: 28 kW
Large focus F2: 34 kW
Focus nominal value Small focus F1: 0.8x0.4 mm
Large focus F2: 0.8x0.7 mm
Anode heat capacity 3.5MHU
Anode target diameter 140mm±0.5mm
Anode target angle 7.0°±0.5°
Weight 8.5kg±0.1kg
Applicable standards GB/T 13797-2009、 GB/T 12078-2012、GB/T 12079-2012、 GB/T 2423.1-2008、GB/T 2423.2-2008、 GB/T 2423.3-2016、 GB/T 2423.10-2019、GB/T 2423.22-2012、 GB/T 191-2008YY/T 0062-2007、 YY/T 0063-2007、YY/T 0064-2016、 YY/T 1099-2007、SJ/T 10624-1995
Safety classification GB 9706.1-2020,GB 9706.228-2020
如需购买服务产品,请与思柯拉特联系
Call Hotline:(+86)18151568323
Address: 4F and 5F, Building No. 1, TC-BIOBAY, No. 52, Yingang Road, Taicang City, Jiangsu Province, China